Powerline™ Catheter manufactured by Biosensors is a rapid exchange catheter designed for percutaneous transluminal coronary angioplasty during the pre-dilation stage. Design Features:
Main element — catheter shaft with a semi-compliant balloon
Two lumens in the shaft:
First — for balloon inflation and deflation
Second — opens approximately 20 cm proximal to the balloon for advancing the coronary guidewire
Technical Specifications:
Maximum coronary guidewire diameter: 0.014"
Nominal balloon inflation pressure: 6 atmospheres
Radiopaque markers at both ends of the balloon (distal and proximal) for accurate positioning
Two additional markers at 90 and 100 cm from the distal tip
Balloon Design Features:
Has the property of diameter increase with increasing pressure
A Luer connector is installed at the proximal end for connection to the inflation device
This design provides high positioning accuracy and efficiency for angioplasty procedures.